• 首页期刊简介编委会刊物订阅专栏专刊电子刊学术动态联系我们English
引用本文:薛巧如,刘冰滢,陈子锋,黄丽华,梁蔚阳,邓锋.ICP-MS与石墨炉原子吸收法测定复方氨基酸注射液中铝含量的比较[J].中国现代应用药学,2021,38(3):318-322.
XUE Qiaoru,LIU Bingying,CHEN Zifeng,HUANG Lihua,LIANG Weiyang,DENG Feng.Comparison of Graphite Furnace Atomic Absorption Spectrometry and ICP-MS for Determination of Aluminum in Compound Amino Acid Injection[J].Chin J Mod Appl Pharm(中国现代应用药学),2021,38(3):318-322.
【打印本页】   【HTML】   【下载PDF全文】   查看/发表评论  【EndNote】   【RefMan】   【BibTex】
←前一篇|后一篇→ 过刊浏览    高级检索
本文已被:浏览 6392次   下载 3138 本文二维码信息
码上扫一扫!
分享到: 微信 更多
ICP-MS与石墨炉原子吸收法测定复方氨基酸注射液中铝含量的比较
薛巧如, 刘冰滢, 陈子锋, 黄丽华, 梁蔚阳, 邓锋
广东省药品检验所, 广州 510663
摘要:
目的 对测定复方氨基酸注射液系列产品中铝含量的电感耦合等离子体质谱(ICP-MS)法与石墨炉原子吸收法(GFAAS)进行比较。方法 采用塞曼偏振背景校正的GFAAS和ICP-MS法对12家企业生产的59批复方氨基酸注射液系列制剂铝离子含量进行测定。结果 GFAAS的铝离子浓度在5~40 ng·mL-1线性关系良好,相关系数(r)为0.995 3,平均回收率为100.1%,RSD为4.80%(n=9);ICP-MS测得的铝离子浓度在5~100 ng·mL-1线性关系良好,相关系数(r)为0.999 9,平均回收率为104.7%,RSD为1.75%(n=9)。结论 2种方法均可用于复方氨基酸注射液制剂中铝离子含量的测定。国内该类制剂的铝离子含量不同生产企业间存在较大差异,有必要控制复方氨基酸注射液中铝的残留量。
关键词:  原子吸收光谱法  电感耦合等离子体质谱    复方氨基酸注射液
DOI:10.13748/j.cnki.issn1007-7693.2021.03.011
分类号:R917.101
基金项目:广州市科技计划项目(201802030002);血液制品质量控制重点实验室(2018ZDB03)
Comparison of Graphite Furnace Atomic Absorption Spectrometry and ICP-MS for Determination of Aluminum in Compound Amino Acid Injection
XUE Qiaoru, LIU Bingying, CHEN Zifeng, HUANG Lihua, LIANG Weiyang, DENG Feng
Guangdong Institute for Drug Control, Guangzhou 510663, China
Abstract:
OBJECTIVE To compare the determination methods for aluminum residual in Compound Amino Acid injection by graphite furnace atomic absorption spectrometry(GFAAS) and inductively coupled plasma-mass spectrometry(ICP-MS). METHODS The content of aluminum in 59 batches of Compound Amino Acid injection produced by 12 pharmaceutical companies was determined by GFAAS with polarized Zeeman background correction and ICP-MS. RESULTS There was a good linear relationship of GFAAS in the concentration range of 5-40 ng·mL-1(r=0.995 3), and the average recovery was 100.1%, RSD=4.80%(n=9). There was a good linear relationship of ICP-MS in the concentration range of 5-100 ng·mL-1 (r=0.999 9), and the average recovery was 104.7%, RSD=1.75%(n=9). CONCLUSION The measured values of the two methods are basically the same. There is a significant difference in residues of aluminum between the products from different manufacturers and it is necessary to control residual aluminum in Compound Amino Acid injection.
Key words:  atomic absorption spectrometry  ICP-MS  aluminum  Compound Amino Acid injection
扫一扫关注本刊微信