• 首页期刊简介编委会刊物订阅专栏专刊电子刊学术动态联系我们English
引用本文:晏海军,陈一波,张福利,林丽娅,王禄.HPLC测定盐酸氯胺酮原料药中的有关物质[J].中国现代应用药学,2022,39(5):633-637.
YAN Haijun,CHEN Yibo,ZHANG Fuli,LIN Liya,WANG Lu.Determination of the Related Substances in Ketamine Hydrochloride by HPLC[J].Chin J Mod Appl Pharm(中国现代应用药学),2022,39(5):633-637.
【打印本页】   【HTML】   【下载PDF全文】   查看/发表评论  【EndNote】   【RefMan】   【BibTex】
←前一篇|后一篇→ 过刊浏览    高级检索
本文已被:浏览 1186次   下载 580 本文二维码信息
码上扫一扫!
分享到: 微信 更多
HPLC测定盐酸氯胺酮原料药中的有关物质
晏海军1,2, 陈一波2, 张福利2, 林丽娅2, 王禄3
1.上海工程技术大学化学化工学院, 上海 201620;2.上海医药工业研究院有限公司, 上海 201203;3.国药集团工业有限公司廊坊分公司, 河北 廊坊 065000
摘要:
目的 建立HPLC检测盐酸氯胺酮原料药中有关物质的方法。方法 采用Waters Xbridge C18色谱柱,以0.005 mol·L-1己烷磺酸钠水溶液-乙腈-磷酸(90︰10︰0.1)为流动相A,乙腈为流动相B,梯度洗脱,流速1.0 mL·min–1,检测波长215 nm,柱温30 ℃。结果 主成分与各杂质间最小分离度>1.5;主成分以及其6个杂质在各自的浓度范围内呈良好的线性关系(r>0.999);平均回收率为97.6%~103.0%,RSD为0.8%~2.8%。结论 本方法准确、灵敏、可靠,适用于盐酸氯胺酮有关物质的测定。
关键词:  盐酸氯胺酮  高效液相色谱法  有关物质
DOI:10.13748/j.cnki.issn1007-7693.2022.05.010
分类号:R917.101
基金项目:
Determination of the Related Substances in Ketamine Hydrochloride by HPLC
YAN Haijun1,2, CHEN Yibo2, ZHANG Fuli2, LIN Liya2, WANG Lu3
1.College of Chemistry and Chemical Engineering, Shanghai University of Engineering Science, Shanghai 201620, China;2.State Pharmaceutical Administration of Shanghai Institute of Pharmaceutical Industry, Shanghai 201203, China;3.China National Pharmaceutical Industry Corporation Ltd., Langfang Company, Langfang 065000, China
Abstract:
OBJECTIVE To establish an HPLC method for determination of the related substances in ketamine hydrochloride.METHODS Waters Xbridge C18 column was used; the mobile phase A consisted of 0.005 mol·L-1 sodium hexanesulfonate aqueous solution-acetonitrile-phosphoric acid (90︰10︰0.1), the mobile phase B was acetonitrile, with gradient elution at the flow rate of 1.0 mL·min-1; the detection wavelength was 215 nm; the column temperature was 30℃. RESULTS The minimum separation between the main component and each impurity was >1.5; ketamine and its six impurities showed a good linear relationship in the self-concentration range (r>0.999); and the average recoveries were 97.6%-103.0% with RSDs of 0.8%-2.8%. CONCLUSION The method is accurate, sensitive and reliable, which is suitable for the determination of related substances in ketamine hydrochloride.
Key words:  ketamine hydrochloride  HPLC  related substance
扫一扫关注本刊微信